Density Measurement of Petroleum Products by Hydrometer Method

Tamson Instruments

 

The Tamson D1298 Density Measurement Apparatus offers precise testing of petroleum products using ASTM methods D287, D891A, D1298, D1429, D1122, IP 160, API 2547, D3142, and D6822, and ISO 3675. It utilizes a Tamson cooling bath to maintain test temperatures within ±0.25°C, crucial for accurate readings. PTFE valves enable quick drainage and cleaning, while a removable spill tray prevents sample spillage. Its design ensures unobstructed hydrometer visibility and scalability with up to six cylinders controlled by a single circulator. This cost-effective solution combines reliability with operational ease for precise density and gravity measurements of petroleum and related products.

What can we do for you?

If you want more specific information regarding this product, please contact our Sales team.
We are happy to assist you.

Features & Benefits

Features:

  • Compliance with Standards: ASTM D287, ASTM D891A, ASTM D1298, ASTM D1429, ASTM D1122, IP 160, API 2547, ASTM D3142, ASTM D6822, ISO 3675.
  • Measurement Capability: Determines density, relative density (specific gravity), or API gravity of petroleum products.
  • Versatility: Suitable for products with Reid vapor pressure of 14.696 psi or less.
  • Temperature Control: Maintains bath temperature within ±0.25°C using Tamson circulator.
  • Construction: Features PTFE valves for efficient drainage, detachable spill tray for clean operation, and unobstructed view of hydrometer scale.
  • Cost Efficiency: Offers competitive pricing compared to thermostatic bath and external chiller combinations.
  • Additional Testing: Capable of ASTM D287 and ASTM D6822 methods.

Benefits:

  • Precision: Ensures sample temperature stability within 0.25°C.
  • Adjustability: Allows fine temperature adjustments (–5.00°C to +5.00°C) post-stabilization.
  • Ambient Compatibility: Operates optimally in ambient temperatures of 18°C to 23°C.
  • Hydrometer Quality: Uses precision hydrometers conforming to BS 718 or DIN 12791.
  • Scalability: Accommodates up to six double wall cylinders with one circulator, expandable with additional racks.

Methods

  • API 2547
  • ASTM D1122
  • ASTM D1298
  • ASTM D1429
  • ASTM D287
  • ASTM D3142
  • ASTM D6822
  • ASTM D891A
  • IP 160
  • ISO 3675